Abstract
This study was performed in order to determine absorbed doses from scattered radiation to personnel during routine equine radiographic procedures and to determine if the protective apparel adequately reduced exposure from scattered radiation. Absorbed doses were measured for one month at the Texas A&M University Veterinary Teaching Hospital using Li:Mg,Cu,P thermoluminescent dosimeters (TLDs). All personnel present in the x-ray examination room during eqine radiography were monitored using TLDs placed at: (1) the finger level; (2) the waist level; (3) the eye level; and (4) the forearm level. Absorbed doses ranged from 0.693 ligy to 31.3 tigy per study. The greatest doses were associated with the individual handling the cassette holder, although the individual making the exposures received similar doses due to improper techniques. The individual holding the horse's halter consistently received the lowest dose. Although all doses observed were within acceptable limits for occupational workers, results demonstrated the need for protective apparel to effectively reduce exposures.
Salinas, Leticia Lamar (1996). Absorbed dose measurements during routine equine radiographic procedures. Master's thesis, Texas A&M University. Available electronically from